  1. every diagnostic imaging system must have __ (2things)
    • protective tube housing
    • correctly functioning control panel
  2. how do we calculate exposure skin entrance dose
    mR/mAs x mAs = mR
  3. the leakage of radiation measured at a distance of 1 m should not exceed ____ when the tube is operated at its highest voltage at the highest current
    100 mR/hr
  4. the carbon fiber material of the radiographic examination table is of how much aluminum equivalency
    is the table radiolucent?
    • 1 mm
    • yes
  5. the indicator must be how accurate of the indicated SID
    2%
  6. what is the benefit of restricting the xray field
    • includes anatomy of clinical interest
    • significant reduction in pt dose because less scatter is produced
    • improves quality xray image
  7. left off at page 3